English German

Saving with SECUTEST is not possible, what can I do?

(Applies to the devices: SECUTEST S2N+ / SIII+)

  • Turn the rotary switch to Menu (S2) and then to Setup or Setup (SIII) and deactivate the Secustore option. There must be a - there.
  • Then go to the Test sequence menu item and deactivate the Direct printing option, which must also contain a -. You can then save the next check.

If it is not possible to deactivate direct printing, please contact our support team