The ID you scanned is not saved?
The ID you scanned is not saved?
(FAQ Nr. PG412)
01 December 2023 - 10:15
|
SECUTEST SII- / SIII-Varianten
|
Database
,
test sequence
Subject:
SECUTEST S2N+ and SIII+
You must make the following setting in the bottom paragraph in the switch positions as follows:
In the S2
:
Line
Passive
Active
EDP
For the SIII in the position Auto respectively for the test sequences:
VDE 0701-0702
EN62353/VDE0751
EN60601
Then proceed as follows:
Select one of these settings and go to SETUP with the down arrow key and then to Database.
There must be a - at the setting ID no. = test sequence.
and for ID no. = 000
At Start with ID-No. you can set an X, but you don't have to.
When going back, save the setting
Do this also with all other settings, then the ID should be stored correctly.