The wrong date and time appears on the log

Important in advance:

  • The date and time must be set correctly in the SECUTEST and in the SI / PSI module.

Further procedure:

  • Importing data from SECUTEST with SECUTEST SI+
  • Click on Data - Import - SECUTEST PSI, not SECUTEST SII or SIII.
daten einlesen 1.png
  • If the reading does not work, check whether the SECUTEST SI+ is correctly entered in the Windows device manager:
daten einlesen 2.jpg
  • If there is no SECUTEST SI+ under Connections (COM & LPT), install the driver Drivercontrol, which you can download from this website: DriverControl

  • A maximum of COM9 can be used. If a higher number was assigned, it must be changed to a number in the range COM1 to COM9. To do this, double-click the SECUTEST SI+ line, go to Port Settings - Advanced and change the COM number accordingly.

Data transfer from older database (Pro Version)

  • Copy the previous database to a USB stick, which you can then plug into a new PC if the program is not installed on the same PC. You can find the file in Elektromanager under Extras - Database.

  • On the new PC click on START (from Windows, but do not start Elektromanager) - Elektromanager Professional and start the DB converter. There you can select the previous database as source and start the converter. Then your data will be transferred to the new installation. After that start Elektromanager and check if the data is available.